A list of known bugs, limitations, and fixed issues for the AtomSpectra ESP32 Gateway.
Status: limitation by design + warning (not a gateway-firmware bug).
The “Atom Spectra” instrument stores its pulse-processing tuning (DSP tuning) inside
itself: pulse-shape thresholds RISE / FALL / NOISE, MAX / HYST / MODE /
STEP, the sampling frequency F, the hardware HV/gain potentiometers POT / POT2,
the pile-up table and the thermal compensation.
Most of these values are visible in the instrument’s reply to the -inf command —
including the pile-up table (PileUp […] / PileUpThr) and the MAX thermal-
compensation table (Tco […]). However, the baseline thermal-compensation table
(POT2 / V versus temperature, the -t_pot points) is NOT returned by -inf —
only the TCpot ON/OFF flag is present there. The full baseline thermal-compensation
snapshot is given by the separate -tc_pot? command.
What was observed. A case was recorded where this entire tuning reset to zero:
-inf started returning RISE 0 FALL 0 NOISE 0 F 1.00 MAX 0 HYST 0 ... POT 0 POT2 0,
while the instrument itself stayed online and still answered VERSION. With zero
thresholds the instrument physically cannot discriminate pulses — counting stops
(total_counts = 0, cps = 0) even though the USB/CDC link is fine. Re-plugging USB
does NOT restore the tuning.
Impact: until the DSP tuning is restored the instrument acquires no spectrum (zero
counts). The gateway, WiFi, Web UI, TCP bridge, and the energy calibration (the
E(ch) polynomial, stored on the gateway in calib.bin) are not affected — this is
a state of the instrument itself, not of the gateway firmware.
⚠ Warning — do NOT restore “blind” with tuning commands.
Although the instrument protocol has configuration commands (-ris, -fall, -nos,
-max, -hyst, -U, -V, -step, etc.), you must not write them at random:
POT / POT2 are hardware HV/gain potentiometers, specific to each individual
detector. Values from another unit (including examples in documentation) can set the
wrong bias/gain and damage the instrument.Workaround / recovery: contact the instrument’s manufacturer (KB Radar) — restoring the factory detector profile is done by the manufacturer. Do not try to write the tuning parameters yourself “blind” (see the warning above).
Recommendation going forward: the gateway can read -inf (it shows the fields in
the Web UI) but does not back them up. While the instrument is correctly tuned it is worth
saving two of its replies once, as a reference “snapshot” of the DSP configuration:
-inf — the main thresholds, POT / POT2, the MAX thermal-
compensation table (Tco […]) and pile-up;-tc_pot? — the baseline thermal-compensation table (POT2 / V
versus temperature); it is not present in -inf (see above).Then a future reset can be detected and both snapshots handed to the manufacturer (KB Radar) as a reference of the factory tuning.
Status: open
The instrument serial number (serial_number) stays empty after connection.
Cause: in response to the -inf command the instrument returns fewer than 40 text lines. The code expects the serial number on line 39 (process_info_response), but the actual response is shorter. The calibration (lines 0–10) is read correctly; the serial number is not.
Impact: in the Web UI and the XML / CSV export the serial number field is empty. Spectrometer functionality is not affected.
Workaround: the serial number can be set manually via the Web UI (calibration panel).
Status: fixed, firmware firmware-v1.0.11.
On the “Monitoring” chart the time-axis unit switch (seconds / minutes / hours) did not change the visible label step — all three positions produced the same grid.
Cause: the old xTickStep (web/monitor.html) computed the target step as
target = spanMs/6, ignoring the selected unit — on a real span (e.g. 199 min) all units
rounded to the same “nice” step (3,600,000 ms), producing an identical grid (all_equal).
Fix: new density-based xTickStep — target tick density is set per unit
(k = 15 / 8 / 4 for s / m / h) over a single nice-list of steps; the rounding guard-wall
was removed. Now s / m / h yield three different steps on any span.
Verified (2026-07-11): on the live board (/api/system fw=v1.0.11) over spans
1 min…12 h the units give distinct=3 different steps; on a 199-min span s=900 s /
m=1800 s / h=3600 s → 13 / 7 / 3 ticks (was: all identical).
Status: fixed, firmware firmware-v1.0.11.
max-width
on the .wrap container. max-width removed, grids switched to auto-fit — the UI now
uses the full window width.Status: implemented, firmware firmware-v1.0.11 (internal format bump on bench build
v1.0.7).
The segmented waterfall format .aswf gained a detector-temperature field (t1 from the
device -inf reply). Format version bumped v4 → v5: header version:5,
row_stride:16406, new temperature field at offset 16402 (float32, °C). When no valid
temperature is available (di->valid = false) a NaN-guard (0x7FC00000) is written.
Compatibility: ASWF v4 readers keep working (the field is appended to the row tail); v5 readers get per-frame temperature.
Status: fixed, firmware firmware-v1.0.11.
sdkconfig — USB-FIFO balance biased toward IN + hostname atomspectra-gw.Status: fixed, firmware firmware-v1.0.6.
Previously “Monitoring” recording (interval-averaged CPS) ran in the browser — with the tab closed or minimized no data accumulated (gaps), and it duplicated the waterfall recording.
Fix: monitoring collection moved onto the board (board-side) — accumulation runs independently of the browser-tab state.
Status: fixed, firmware firmware-v1.0.6.
When stitching .aswf segments into a single waterfall, stripes / time unevenness appeared
at the boundaries.
Cause: the segment header wrote row_stride as 16406 instead of the actual 16402 —
a row-step desync during stitching (main/web_waterfall.c).
Fix: header row_stride corrected to the actual 16402. Files captured by the old
firmware are repaired by recomputing the step (PC-side reconciliation utility).
Status: implemented, firmware firmware-v1.0.6.
#N · −Mm Ss, live → LIVE).-inf/-cal reply race in bridge mode corrupted serial and temperatureStatus: fixed, firmware firmware-v1.0.5.
In TCP-bridge mode (port 8234), polling the device concurrently with -inf and -cal
mixed the replies: the serial number degenerated to FFFFFFFF, temperature read as 0,
and fields from different replies fused together.
Fix: separation and serialization of the -inf and -cal reply streams in the bridge
path.
Verified (#TEST-2 Block A, 2026-07-09): 40 iterations of interleaved -inf/-cal over
TCP :8234 — 0 corrupted serial / t1 / version.
Status: implemented, firmware firmware-v1.0.4.
data_cb → RX ring → usb_rxw); an rx_ring_drops metric was added
to the status JSON..aswf row + a global segment seq (NVS-persisted, gap detection) + a reconciliation
counter “device-total vs written”. The PC client verifies CRC/seq during assembly.Verified (#TEST-2 Block A): rx_ring_drops delta = 0; per-row CRC 64/64; n42 ↔ aswf
192/192.
Status: implemented, firmware firmware-v1.0.4.
The “Monitoring” chart gained a time X-axis (seconds / minutes / hours + real time); the timezone is set on the “System” page.
Status: implemented, firmware firmware-v1.0.4 (bench build v1.0.3).
StartDateTime 1970 on every row when auto-starting after rebootStatus: fixed in 79eff24, in main. Firmware firmware-v1.0.1.
When waterfall recording auto-started after a reboot/power-up, the n42 export showed
<StartDateTime>1970-01-01T...Z</StartDateTime> on every segment — the start time was not
synchronized with the real clock.
Cause: initialization order in app_main() — usb_host_cdc_init() (waterfall
boot-autostart) runs before init_sntp(). spectrogram_start() latches
started_at = time(NULL) while the RTC is still at epoch 0 (no SNTP reply yet), so
started_at is pinned near 1970. Reconnecting USB/WiFi does not fix the value.
Fix: added an SNTP callback spectrogram_time_synced() (main/spectrogram.c:831).
On the first SNTP reply, if recording is active and started_at < WF_SANE_EPOCH,
started_at is recomputed backwards from elapsed time: started_at = time(NULL) − elapsed.
The real recording start is restored retroactively without losing already-written segments.
Verification (2026-07-08): a field n42 export waterfall (30).n42 pulled from the board
(firmware firmware-v1.0.1, flashed with the flasher) — 60 segments, zero 1970, all
dated 2026-07-08; the boot-autostart segment is dated correctly (12:24:51Z), and
StartDateTime increases monotonically. Fixed on hardware.
/ws/waterfall returns 404 → endless reconnect loopStatus: fixed in 6c81c37, in main.
The waterfall WebSocket endpoint /ws/waterfall intermittently returned 404 Not Found
and the browser fell into a reconnect loop — the waterfall stream never came up.
Cause: the ESP-IDF httpd URI-handler registry overflowed. CONFIG_HTTPD_MAX_URI_HANDLERS
was 45, while the total number of registered handlers (pages, assets, REST API, segment
endpoints) grew past 45. Registration of /ws/waterfall failed with
ESP_ERR_HTTPD_HANDLERS_FULL → the server had no route → 404.
Fix: CONFIG_HTTPD_MAX_URI_HANDLERS raised 45 → 60 in sdkconfig.defaults — headroom
for current and future endpoints.
Status: fixed in 33fb4a4 (+ 2b36737), in main.
Under load from active waterfall recording (writing .aswf segments + simultaneous USB
intake from the detector + WiFi polling), the device crash-looped — rebooting roughly
every ~15 minutes.
Cause: CONFIG_SPI_FLASH_AUTO_SUSPEND was enabled on a flash chip not on the
ESP-IDF whitelist for that option — under concurrent load the SPI flash entered
auto-suspend at an inopportune moment during a write, causing a crash.
Fix: CONFIG_SPI_FLASH_AUTO_SUSPEND was disabled.
Verification (#STAB-2, 2026-07-04): a 9.40 h continuous board-path waterfall
recording retest (not an emulation) — 0 reboots, 0 dropped flash segments
(seg_dropped) over the whole run, 52 clean SEG_ROLLOVER events. Before the fix — a
reboot every ~15 min under the same load. Full report with telemetry and charts:
docs/stab2_report.md (web version with charts).
spectrogram_is_recordingStatus: fixed in 1b21d61
The GitHub Actions build job (ESP-IDF v5.4) failed on a clean clone of the repository
within ~2 minutes. Local builds passed — the bug only showed up on a fresh clone (CI, a
different machine).
Cause: an unfinished fragment of the USB reconnect logic leaked into the public
repository. In main/usb_host_cdc.c (after the CDC link to the device was restored)
there was a call to spectrogram_is_recording() — to resume spectrum acquisition with
the -sta command if waterfall recording was active. That function is defined in the
waterfall recording subsystem files, which were not committed at the time of that
commit (the feature was unfinished local work). On the author’s machine everything
compiled because the definition was in the working copy; on a clean clone the compiler
saw only the call with no declaration → implicit declaration of function
'spectrogram_is_recording'. ESP-IDF builds with -Werror=all, so an implicit
declaration is a hard build error, not a warning.
Fix: the unfinished recording-resume hook was removed from usb_host_cdc.c in the
public repository — the build again depends only on committed symbols. The
auto-resume-on-reconnect feature itself stays in an unpublished local branch and will
land as a single coherent commit together with the whole stream-to-disk recording
subsystem (not as a lone dangling call).
Update: the promised single commit has landed — the autonomous segment-recording
subsystem (spectrogram.[ch], the /segments and /segment endpoints, the reconnect
hook in usb_host_cdc.c, auto-resume via spectrogram_restore()) is published as a
whole in the rec-11-autonomous-recording branch. A clean-clone build is self-contained
again — spectrogram_is_recording() is now defined within the same commit as its call.
Verification: the CI build job on commit 1b21d61 is green (the prior b0b8856
and 958ac19 failed with the same implicit declaration). Prevention going forward:
before pushing, verify the branch “as a clean clone” — build from committed files only,
not from a working copy that has uncommitted changes.
Status: fixed (see web/waterfall.html, write serializer dskEnqueue)
While recording the waterfall to a file (“💾 Stream to disk”, .aswf format via the
browser File System Access API) the recording would stop on its own after a while —
the recording indicator went off and the file stopped growing, even though the
WebSocket stream kept arriving. It reproduced more reliably the more frequently rows
arrived (smaller waterfall interval); at larger intervals it happened less often but
still systematically.
Cause: incoming WS rows were written to the FileSystemWritableFileStream with
dskWritable.write(...) without await, while every 30 seconds the
flushDiskHeader timer rewrote the file header (several more write() calls).
FileSystemWritableFileStream.write() is asynchronous: starting a new write while a
previous one is still pending throws synchronously (the stream is locked,
The stream is locked). At the “data-row write ↔ header rewrite” boundary the
exception occurred systematically (roughly every 30 seconds); it was caught by the
write error handler, which called stopDiskStream() — so the recording disabled
itself.
Fix: all writes to the stream are serialized through a single Promise queue
(dskEnqueue) — the next write starts only after the previous one completes, never
two concurrent write() calls. Write positions use explicit byte offsets
({type:"write",position:…}) instead of the implicit cursor: the header (positions 0
and 8) and the data rows (position ≥ 8 + header reserve) no longer overlap.
stopDiskStream() drains the queue (await dskQueue) before the final header write
and close(); a reentrancy guard dskStopping was added.
Verification: firmware build (ESP-IDF v5.4) with no errors; node --check of the
embedded JS — no syntax errors; the change is mirrored into the demo
(demo/waterfall.html). On-device functional test: waterfall recording at a 5 s
interval (rows arrive more often than the 30 s header flush fires) — the recording
keeps running without self-disabling.
Status: fixed (see sdkconfig.defaults, key CONFIG_HTTPD_MAX_REQ_HDR_LEN)
On repeated hard reload (Ctrl+F5) of the waterfall page the UI could break — some assets failed to load, charts/stream stopped updating.
Cause: on a hard reload the browser bypasses the cache and sends a request with
the full set of HTTP headers (Authorization: Basic from web-auth + Sec-Fetch-* +
Accept* + Cache-Control: no-cache). The combined header block exceeded the
ESP-IDF httpd limit CONFIG_HTTPD_MAX_REQ_HDR_LEN (default 512 B), so the server
replied 431 Request Header Fields Too Large instead of the page. The firmware stayed
fully stable throughout (heap/CPU unchanged, no reboots) — only the browser UI broke.
Fix: httpd limits raised in sdkconfig.defaults:
CONFIG_HTTPD_MAX_REQ_HDR_LEN=1024 and CONFIG_HTTPD_MAX_URI_LEN=1024 (512 → 1024 B).
Verification: Ctrl+F5 stress test with UART capture — after the fix the 431
responses are gone (was: on every F5 → now: 0), no firmware reboots, WS reconnects
succeed. Method and raw before/after logs: tests/stress/.
Status: fixed in 3ab490f
After an ESP32 reboot, calibration could be missing from the JSON API and XML export despite a saved calib.bin.
Cause: in main.c the call to spectrum_load_calibration() came before spectrum_restore_autosave(). The autosave restored the full s_spectrum structure via fread(), overwriting the calibration loaded from calib.bin.
Fix: the call order in app_main() was changed — first spectrum_restore_autosave(), then spectrum_load_calibration().
Status: fixed in a9dab5e
When hovering the cursor over the spectrum, the keV value did not match the real channel energy.
Cause: in the Web UI the energy calculation applied the polynomial coefficients incorrectly — the coefficient order was inverted (the BecqMoni UI shows coefficients in descending order: a=ch⁴, b=ch³, …, e=offset, while the internal format is ascending: c₀=offset, c₁=ch¹, …).
Fix: the channelToKeV() function in index.html recomputes using the ascending polynomial E(ch) = c₀ + c₁·ch + c₂·ch² + c₃·ch³ + c₄·ch⁴.
Status: fixed in a9dab5e
Fix: a full Web UI rework — protected Math.log10() against zeros, correct device name.
Status: fixed in 7cf0beb
While manually entering calibration coefficients via the Web UI, the update() loop (once per second) overwrote the field contents with the current values from the API.
Fix: a calibEditing flag was added — while the user edits the form, auto-refresh of the calibration fields is paused.
Autosave of the current spectrum (current.bin, ~33 KB) runs every 60 seconds. That is ~1440 writes per day.
| Flash type | Endurance | Estimated lifetime |
|---|---|---|
| Winbond (typical, 100K cycles) | 100,000 erases/sector | 15–30 years |
| Cheap no-name (10K cycles) | 10,000 erases/sector | 1.5–5 years |
LittleFS uses copy-on-write and wear leveling across the whole partition (12.9 MB), which spreads the load. With original ESP32-S3-DevKitC-1 boards using Winbond flash, no issues are expected.
Possible optimizations (not implemented, not needed at current lifetimes):
The TCP bridge (port 8234) supports one simultaneous connection. A second connection attempt is rejected. BecqMoni or AtomSpectra on a PC works until a second instance opens.
5 GHz networks are not supported in hardware. Make sure the router broadcasts on 2.4 GHz.
The Atom Spectra instrument transmits 8192 channels. This is a hardware limitation of the spectrometer, not the firmware.
Context: the feature + the #WF-1 fix (33fb4a4) are already in main; the
rec-11-autonomous-recording branch as a whole is not merged yet (extra unrelated commits on top).
Autonomous segment recording (seg_NNNNN.aswf) and segment rotation when
WF_SEG_MAX_ROWS (64 rows) is reached are confirmed on the device: seg_00000.aswf
finalizes when full and seg_00001.aswf is created automatically.
The keep-last ring mode (on flash-full the oldest not-yet-offloaded segment is
overwritten; counters seg_count / seg_dropped) is implemented but has not been
verified by a long soak up to actual partition exhaustion (≈763 rows, ≈5 days at a
10-min interval). Until that test passes, the “flash full → overwrite oldest” boundary is
considered untested — that specific scenario (real partition exhaustion) hasn’t been run yet,
though the recording code itself is already in main.
Update 2026-07-04 (#STAB-2): a 9.40 h board-path soak test confirmed the reliability
of the segment writes themselves — 0 reboots, 0 seg_dropped, 52 clean
SEG_ROLLOVER events (see the #WF-1 fix above). This is not the same test as a soak to
actual exhaustion of the 763-row partition (the segment keep-last ring still has not been
run to real overflow) — but a separate export limitation was found along the way: the
ring_capacity field in /api/waterfall/status is fixed at 256 rows (~4.25 h at a
~60 s cadence) — smaller than the ~763-row partition-capacity estimate above. n42 exports
of recordings longer than ~4.25 h come out truncated (see #FW-19 below). Full writeup:
docs/stab2_report.md §6.
Status: open (task #32).
When exporting to n42 a recording that ran longer than ~4.25 h, the file contains only
the last 256 rows — regardless of how many rows were actually recorded
(total_rows). Cause: ring_capacity=256 (/api/waterfall/status) — the ring
physically holds only the last 256 rows; older ones get overwritten by new ones before
the export is pulled.
This is not a #WF-1 regression — the seg_dropped counter (write losses) stays at 0,
i.e. the write path itself loses no data; the limitation is in the exported ring’s
storage size, not write reliability. The RING_OVERFLOW event does not detect the
overwrite itself (its condition is the write-lag total_rows − flash_rows ≥
ring_capacity, not the wraparound fact).
Workaround: for recordings longer than ~4.25 h, periodically pull segments via
/api/waterfall/segment (see WATERFALL.en.md)
instead of waiting until the end of the recording for a single export. Details and the
telemetry from the test that found this: docs/stab2_report.md §6.
wf_pull_client.py, #REC-12): no pin against the keep-last ringThe keep-last ring (make_room(), main/spectrogram.c:304-321) deletes the oldest
FINALIZED segment when Flash space is short — except the currently open segment and any
pinned one (s_seg_pinned). Only the push offload path (#REC-11-A2, wf_offload.c)
pins a segment; the pull path (GET /api/waterfall/segment, main/web_waterfall.c:468)
does not pin the segment while it downloads it.
If the PC client polls SLOWER than the board fills Flash with unread segments, the ring
can delete a segment before wf_pull_client.py gets a chance to fetch it. The next
GET .../segment?name=... then fails (error:get:... in the client log), no delete/ack
is sent, but the segment is already physically gone — the data is lost for good, leaving
a hole in the stitched .aswf. This is NOT detected as a “board pause” (the gap check
in wf_pull_client.py:161-169 only catches recording pauses, not segments the ring ate).
Workaround: keep the client’s --interval well below the time it takes the ring to
eat an unread segment at the current recording rate. There is currently no automatic
protection (a pin, like push has) for the pull download path — not implemented.
| Component | Version |
|---|---|
| ESP-IDF | v5.1+ (tested on v5.4) |
| Target chip | ESP32-S3 (USB OTG Host required) |
| Spectrometer | KB Radar “Atom Spectra” (FTDI FT232R, 600000 baud) |
| BecqMoni XML | FormatVersion 120920 |
| Web UI | Modern browsers (Chrome, Firefox, Safari, Edge) |